view Documentations/dump-rte.cfg @ 228:f9ba924d188e div-fixes-4-1

Bugfix: set battery percentage correctly after RTE update When the RTE is rebooted, the battery percentage on the display is zeroed. However, when selecting SYS2-reboot-maintenance, the previously known battery percentage can be manually restored. Interestingly, when the restore reported a percentage A, choosing that resulted in a percentage B to be displayed again. With B much smaller than A. So, rebooting the RTE multiple times resulted in an seemingly empty battery, while it definitely is not. The reason for this is a relatively simple bug in the RTE code. This commit fixes the conversion between the internal LTC2941 registers and the percentage value to be displayed. Obviously, from and to the internal registers need to be symmetrical. Signed-off-by: Jan Mulder <jlmulder@xs4all.nl>
author Jan Mulder <jlmulder@xs4all.nl>
date Tue, 02 Apr 2019 14:53:15 +0200
parents 01cc5959f199
children
line wrap: on
line source

#! openocd -f
# Define the prob used:
source [find interface/stlink-v2.cfg]
set WORKAREASIZE 0x8000
transport select hla_swd

# Reset options
set ENABLE_LOW_POWER 1
set STOP_WATCHDOG 1
reset_config srst_only srst_nogate connect_assert_srst

# Seelct the right chip
set CHIPNAME stm32f411RETx
set CONNECT_UNDER_RESET 1
source [find target/stm32f4x.cfg]

# Allow to continue execution after a connection:
init_reset run

#puts "Flash banks:"
#flash banks

#puts "Reading..."
#flash dump_image CPU2-RTE-dump.hex 0x00000000 0x8000

#puts "Done."
#exit