view Small_CPU/Inc/uart_Internal.h @ 977:f0622dd93fdf Evo_2_23

Bugfix ppo2 display in simulator mode: In the previous version a wrong ppo2 value was display if a digital O2 sensor was used. Rootcause was that the calibration data of the analog sensors was used. To fix this a default calibration coefficient is used (calibration is done within the sensor => no need to store it in OSTC) if the sensor is identified as digital O2 sensor.
author Ideenmodellierer
date Thu, 30 Jan 2025 21:24:46 +0100
parents effadaa3a1f7
children
line wrap: on
line source

/**
  ******************************************************************************
  * @file    uartInternal.h
  * @author  heinrichs weikamp gmbh
  * @version V0.0.1
  * @date    03-November-2024
  * @brief   button control
  *           
  ******************************************************************************
  * @attention
  *
  * <h2><center>&copy; COPYRIGHT(c) 2015 heinrichs weikamp</center></h2>
  *
  ******************************************************************************
  */ 

/* Define to prevent recursive inclusion -------------------------------------*/
#ifndef UARTINTERNAL_H
#define UARTINTERNAL_H

#ifdef __cplusplus
 extern "C" {
#endif

#include "stm32f4xx_hal.h"

UART_HandleTypeDef huart6;

void MX_USART6_UART_Init(void);
void MX_USART6_DMA_Init(void);
void MX_USART6_UART_DeInit(void);
void GNSS_IO_init(void);

void UART6_HandleUART();

#ifdef __cplusplus
}
#endif

#endif /* UARTINTERNAL_H */

/************************ (C) COPYRIGHT heinrichs weikamp *****END OF FILE****/