comparison Discovery/Src/externLogbookFlash.c @ 322:31e471d60797 O2_SensorSync

Added start of frame detection for HUD data In the previous impmenentation the reception and evaluation of 15 byte were used without start detection. As a result sensor data could be stuck in case the frame sequence does not match the structure (e.g. cause by an framing error). To resolve this in case of an invalid checksumme in combination with a data lost detection a byte based reception is startet to detect the start of a frame using the break between two transmissions. In addition a babbling idiot protecting has been added because a faulty hardware could have an impact on the OSTC operation (high interrupt load)
author ideenmodellierer
date Sun, 30 Jun 2019 21:25:58 +0200
parents c7e665e0b08f
children 2174fb133dbe
comparison
equal deleted inserted replaced
321:37ee61f93124 322:31e471d60797